      Comparison of Binax Legionella Urinary Antigen EIA kit with Binax RIA Urinary Antigen kit for detection of Legionella pneumophila serogroup 1 antigen.

      1 , , , ,
      Journal of clinical microbiology

      Read this article at

      ScienceOpenPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The Legionella Urinary Antigen EIA kit (Binax, Portland, Maine) was compared with the EQUATE RIA Legionella Urinary Antigen kit (Binax) for its ability to detect the presence of urinary antigens to Legionella pneumophila serogroup 1. Urine specimens from patients without Legionnaires' disease (n = 33) were negative by both methods (specificity, 100%). Twenty (77%) of 26 urine specimens from patients with Legionnaires' disease positive by the radioimmunoassay kit were also positive by the enzyme immunoassay (EIA) kit. If the cutoff for a positive EIA result were lowered to a ration of > or = 2.5, 23 of 26 (88%) urine specimens would have been positive by EIA and the specificity would remain 100%. Use of the EIA kit is an acceptable method for detecting L. pneumophila serogroup 1 urinary antigens by laboratories that do not want to handle radioactive materials.

          Related collections

          Author and article information

          Journal
          J. Clin. Microbiol.
          Journal of clinical microbiology
          0095-1137
          0095-1137
          Jun 1996
          : 34
          : 6
          Affiliations
          [1 ] Ohio State University Medical Center, Columbus 43210, USA.
          Article
          229069
          8735125
          93fb9aa9-294d-45a4-93e6-fc2ed58c99f6
          History

          Comments

          Comment on this article