      Optic nerve head parameters of high-definition optical coherence tomography and Heidelberg retina tomogram in perimetric and preperimetric glaucoma

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background:

          Heidelberg retina tomogram (HRT) and optical coherence tomography (OCT) are two widely used imaging modalities to evaluate the optic nerve head (ONH) in glaucoma.

          Purpose:

          To compare the ONH parameters of HRT3 and high-definition OCT (HD-OCT) and evaluate their diagnostic abilities in perimetric and preperimetric glaucoma.

          Design:

          Cross-sectional analysis.

          Methods:

          35 control eyes (24 subjects), 21 preperimetric glaucoma eyes (15 patients), and 64 perimetric glaucoma eyes (44 patients) from the Longitudinal Glaucoma Evaluation Study underwent HRT3 and HD-OCT examinations.

          Statistical Analysis:

          Agreement between the ONH parameters of HRT and HD-OCT were assessed using Bland-Altman plots. Diagnostic abilities of ONH parameters were evaluated using area under the receiver operating characteristic curves (AUCs), sensitivity at fixed specificity, and likelihood ratios (LR).

          Results:

          Optic disc area, vertical cup to disc ratio, and cup volume with HD-OCT were larger than with HRT, while the rim area was smaller with HD-OCT ( P < 0.001 for all comparisons). AUCs of all HD-OCT ONH parameters (0.90-0.97 in perimetric and 0.62-0.71 in preperimetric glaucoma) were comparable ( P > 0.10) to the corresponding HRT ONH parameters (0.81-0.95 in perimetric and 0.55-0.72 in preperimetric glaucoma). LRs associated with diagnostic categorization of ONH parameters of both HD-OCT and HRT were associated with larger effects on posttest probability of perimetric compared to preperimetric glaucoma.

          Conclusions:

          ONH measurements of HD-OCT and HRT3 cannot be used interchangeably. Though the diagnostic abilities of ONH parameters of HD-OCT and HRT in glaucoma were comparable, the same were significantly lower in preperimetric compared to perimetric glaucoma.

          Related collections

          Most cited references49

          • Record: found
          • Abstract: found
          • Article: not found

          Measuring agreement in method comparison studies.

          Agreement between two methods of clinical measurement can be quantified using the differences between observations made using the two methods on the same subjects. The 95% limits of agreement, estimated by mean difference +/- 1.96 standard deviation of the differences, provide an interval within which 95% of differences between measurements by the two methods are expected to lie. We describe how graphical methods can be used to investigate the assumptions of the method and we also give confidence intervals. We extend the basic approach to data where there is a relationship between difference and magnitude, both with a simple logarithmic transformation approach and a new, more general, regression approach. We discuss the importance of the repeatability of each method separately and compare an estimate of this to the limits of agreement. We extend the limits of agreement approach to data with repeated measurements, proposing new estimates for equal numbers of replicates by each method on each subject, for unequal numbers of replicates, and for replicated data collected in pairs, where the underlying value of the quantity being measured is changing. Finally, we describe a nonparametric approach to comparing methods.
            Bookmark
            • Record: found
            • Abstract: not found
            • Book: not found

            The statistical evaluation of medical tests for classification and prediction

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Ability of cirrus HD-OCT optic nerve head parameters to discriminate normal from glaucomatous eyes.

              To determine the ability of optic nerve head (ONH) parameters measured with spectral domain Cirrus HD-OCT (Carl Zeiss Meditec, Inc., Dublin, CA) to discriminate between normal and glaucomatous eyes and to compare them with the discriminating ability of peripapillary retinal nerve fiber layer (RNFL) thickness measurements performed with Cirrus HD-OCT. Evaluation of diagnostic test or technology. Seventy-three subjects with glaucoma and 146 age-matched normal subjects. Peripapillary ONH parameters and RNFL thickness were measured in 1 randomly selected eye of each participant within a 200 × 200 pixel A-scan acquired with Cirrus HD-OCT centered on the ONH. Optic nerve head topographic parameters, peripapillary RNFL thickness, and area under receiver operating characteristic curves (AUCs). To distinguish normal from glaucomatous eyes, regardless of disease stage, the 6 best parameters (expressed as AUC) were vertical rim thickness (VRT, 0.963), rim area (0.962), RNFL thickness at clock-hour 7 (0.957), RNFL thickness of the inferior quadrant (0.953), vertical cup-to-disc ratio (VCDR, 0.951), and average RNFL thickness (0.950). The AUC for distinguishing between normal eyes and eyes with mild glaucoma was greatest for RNFL thickness of clock-hour 7 (0.918), VRT (0.914), rim area (0.912), RNFL thickness of inferior quadrant (0.895), average RNFL thickness (0.893), and VCDR (0.890). There were no statistically significant differences between AUCs for the best ONH parameters and RNFL thickness measurements (P > 0.05). Cirrus HD-OCT ONH parameters are able to discriminate between normal eyes and eyes with glaucoma or even mild glaucoma. There is no difference in the ability of ONH parameters and RNFL thickness measurement, as measured with Cirrus OCT, to distinguish between normal and glaucomatous eyes. Copyright © 2011 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.
                Bookmark

                Author and article information

                Journal
                Indian J Ophthalmol
                Indian J Ophthalmol
                IJO
                Indian Journal of Ophthalmology
                Medknow Publications & Media Pvt Ltd (India )
                0301-4738
                1998-3689
                April 2016
                : 64
                : 4
                : 277-284
                Affiliations
                [1]VST Glaucoma Center, L. V. Prasad Eye Institute, Hyderabad, Telangana, India
                Author notes
                Correspondence to: Dr. Harsha Laxmana Rao, L. V. Prasad Eye Institute, Kallam Anji Reddy Campus, Banjara Hills, Hyderabad - 500 034, Telangana, India. E-mail: harshalaxmanarao@ 123456gmail.com
                Article
                IJO-64-277
                10.4103/0301-4738.182938
                4901845
                27221679
                49eb2004-d12a-4b2b-9b3e-0a067585cbe0
                Copyright: © Indian Journal of Ophthalmology

                This is an open access article distributed under the terms of the Creative Commons Attribution NonCommercial ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non commercially, as long as the author is credited and the new creations are licensed under the identical terms.

                History
                : 14 April 2015
                : 09 February 2016
                Categories
                Original Article

                Ophthalmology & Optometry
                glaucoma,heidelberg retina tomogram,optic nerve head,optical coherence tomography

                Comments

                Comment on this article

                scite_
                0
                0
                0
                0
                Smart Citations
                0
                0
                0
                0
                Citing PublicationsSupportingMentioningContrasting
                View Citations

                See how this article has been cited at scite.ai

                scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

                Similar content194

                Cited by3

                Most referenced authors518