26
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Development and long‐term stability of a comprehensive daily QA program for a modern pencil beam scanning (PBS) proton therapy delivery system

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          The main purpose of this study is to demonstrate the clinical implementation of a comprehensive pencil beam scanning ( PBS) daily quality assurance ( QA) program involving a number of novel QA devices including the Sphinx/Lynx/parallel‐plate ( PPC05) ion chamber and HexaCheck/multiple imaging modality isocentricity ( MIMI) imaging phantoms. Additionally, the study highlights the importance of testing the connectivity among oncology information system (OIS), beam delivery/imaging systems, and patient position system at a proton center with multi‐vendor equipment and software.

          Methods

          For dosimetry, a daily QA plan with spot map of four different energies (106, 145, 172, and 221 MeV) is delivered on the delivery system through the OIS. The delivery assesses the dose output, field homogeneity, beam coincidence, beam energy, width, distal‐fall‐off (DFO), and spot characteristics — for example, position, size, and skewness. As a part of mechanical and imaging QA, a treatment plan with the MIMI phantom serving as the patient is transferred from OIS to imaging system. The HexaCheck/ MIMI phantoms are used to assess daily laser accuracy, imaging isocenter accuracy, image registration accuracy, and six‐dimensional (6D) positional correction accuracy for the kV imaging system and robotic couch.

          Results

          The daily QA results presented herein are based on 202 daily sets of measurements over a period of 10 months. Total time to perform daily QA tasks at our center is under 30 min. The relative difference (Δ rel) of daily measurements with respect to baseline was within ± 1% for field homogeneity, ±0.5 mm for range, width and DFO, ±1 mm for spots positions, ±10% for in‐air spot sigma, ±0.5 spot skewness, and ±1 mm for beam coincidence (except 1 case: Δ rel = 1.3 mm). The average Δ rel in dose output was −0.2% (range: −1.1% to 1.5%). For 6D IGRT QA, the average absolute difference (Δ abs) was ≤0.6 ± 0.4 mm for translational and ≤0.5° for rotational shifts.

          Conclusion

          The use of novel QA devices such as the Sphinx in conjunction with the Lynx, PPC05 ion chamber, HexaCheck/ MIMI phantoms, and my QA software was shown to provide a comprehensive and efficient method for performing daily QA of a number of system parameters for a modern proton PBS‐dedicated treatment delivery unit.

          Related collections

          Most cited references15

          • Record: found
          • Abstract: found
          • Article: not found

          Task Group 142 report: quality assurance of medical accelerators.

          The task group (TG) for quality assurance of medical accelerators was constituted by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance and Outcome Improvement Subcommittee. The task group (TG-142) had two main charges. First to update, as needed, recommendations of Table II of the AAPM TG-40 report on quality assurance and second, to add recommendations for asymmetric jaws, multileaf collimation (MLC), and dynamic/virtual wedges. The TG accomplished the update to TG-40, specifying new test and tolerances, and has added recommendations for not only the new ancillary delivery technologies but also for imaging devices that are part of the linear accelerator. The imaging devices include x-ray imaging, photon portal imaging, and cone-beam CT. The TG report was designed to account for the types of treatments delivered with the particular machine. For example, machines that are used for radiosurgery treatments or intensity-modulated radiotherapy (IMRT) require different tests and/or tolerances. There are specific recommendations for MLC quality assurance for machines performing IMRT. The report also gives recommendations as to action levels for the physicists to implement particular actions, whether they are inspection, scheduled action, or immediate and corrective action. The report is geared to be flexible for the physicist to customize the QA program depending on clinical utility. There are specific tables according to daily, monthly, and annual reviews, along with unique tables for wedge systems, MLC, and imaging checks. The report also gives specific recommendations regarding setup of a QA program by the physicist in regards to building a QA team, establishing procedures, training of personnel, documentation, and end-to-end system checks. The tabulated items of this report have been considerably expanded as compared with the original TG-40 report and the recommended tolerances accommodate differences in the intended use of the machine functionality (non-IMRT, IMRT, and stereotactic delivery).
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Impact of Spot Size and Beam-Shaping Devices on the Treatment Plan Quality for Pencil Beam Scanning Proton Therapy.

            This study aimed to assess the clinical impact of spot size and the addition of apertures and range compensators on the treatment quality of pencil beam scanning (PBS) proton therapy and to define when PBS could improve on passive scattering proton therapy (PSPT).
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of prostate proton treatment planning technique, interfraction robustness, and analysis of single-field treatment feasibility

              This study compares target coverage robustness among proton therapy plans for prostate cancer patients treated with 2 laterally opposed fields delivered daily or, alternatively, every other day as single lateral fields, using uniform scanning (US), single-field uniform dose (SFUD), pencil beam scanning (PBS) optimized for uniform target coverage only, SFUD PBS optimized for target coverage and organs at risk (OAR) sparing (SFUD-opt), and intensity modulated proton therapy (IMPT).
                Bookmark

                Author and article information

                Contributors
                suresh.rana@gmail.com
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                28 March 2019
                April 2019
                : 20
                : 4 ( doiID: 10.1002/acm2.2019.20.issue-4 )
                : 29-44
                Affiliations
                [ 1 ] Department of Radiation Oncology Miami Cancer Institute Baptist Health South Florida Miami FL USA
                [ 2 ] Department of Radiation Oncology Herbert Wertheim College of Medicine Florida International University Miami FL USA
                [ 3 ] Department of Physics School of Advanced Sciences VIT University Vellore India
                Author notes
                [*] [* ] Author to whom correspondence should be addressed. Suresh Rana

                E‐mail: suresh.rana@ 123456gmail.com ; Telephone: 405‐795‐6697.

                Article
                ACM212556
                10.1002/acm2.12556
                6448164
                30920146
                fde77d4c-ea0a-409d-87ab-ee805cdbc3a1
                © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

                This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 02 October 2018
                : 16 December 2018
                : 22 January 2019
                Page count
                Figures: 14, Tables: 4, Pages: 16, Words: 8156
                Categories
                87.55.Qr
                87.56.bd
                87.56.Fc
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm212556
                April 2019
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.6.2.1 mode:remove_FC converted:04.04.2019

                daily qa,pencil beam scanning,proton therapy
                daily qa, pencil beam scanning, proton therapy

                Comments

                Comment on this article