          Abstract

          Background

          Wound healing is a goal for advanced technology in the surgical space to benefit clinical outcomes. Surgical staplers are commonly used in a variety of open and minimally invasive abdominal and thoracic procedures. Assessment of wound healing traits, such as perfusion, has been challenging due to technical limitations. A novel technique that utilizes micro-computed tomography methodology to measure perfusion was designed to compare the micro-perfusion of staple lines between commercial stapler reloads that employ different staple height strategies.

          Materials and methods

          Following an Institutional Animal Care and Use Committee-approved protocol, rats were euthanized and immediately heparinized prior to a subtotal gastrectomy with either graduated-height or single-height staples. Rats were then perfused with barium, following which stomachs were removed and immediately fixed in formalin to prevent degradation. Stomachs were then imaged using micro-computed tomography and subsequent analysis was utilized to quantify fluid volume and patent vasculature proximity to staples within the staple line region for each group.

          Results

          Average perfusion volume was significantly higher with graduated-height staples (0.33% ± 0.18%) compared to single-height staples (0.16% ± 0.09%, P=0.011). Average vessel-to-staple line distance was not significant but trended lower with graduated-height staples (0.35±0.02 mm) compared to single-height staples (0.36±0.03 mm, P=0.18).

          Discussion

          Graduated-height staples had significantly higher perfusion volume than single-height staples, which likely has a downstream benefit on wound healing and clinical outcomes.

          Conclusion

          This study shows a higher perfusion volume around the staple lines using graduated-height staples as compared to single-height staples and this may contribute to better wound healing in patients.

          Related collections

          Most cited references18

          • Record: found
          • Abstract: found
          • Article: not found

          Oxygen in acute and chronic wound healing.

          Oxygen is a prerequisite for successful wound healing due to the increased demand for reparative processes such as cell proliferation, bacterial defence, angiogenesis and collagen synthesis. Even though the role of oxygen in wound healing is not yet completely understood, many experimental and clinical observations have shown wound healing to be impaired under hypoxia. This article provides an overview on the role of oxygen in wound healing and chronic wound pathogenesis, a brief insight into systemic and topical oxygen treatment, and a discussion of the role of wound tissue oximetry. Thus, the aim is to improve the understanding of the role of oxygen in wound healing and to advance our management of wound patients.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Sensors and imaging for wound healing: a review.

            Wound healing involves a complex series of biochemical events and has traditionally been managed with 'low tech' dressings and bandages. The concept that diagnostic and theranostic sensors can complement wound management is rapidly growing in popularity as there is tremendous potential to apply this technology to both acute and chronic wounds. Benefits in sensing the wound environment include reduction of hospitalization time, prevention of amputations and better understanding of the processes which impair healing. This review discusses the state-of-the-art in detection of markers associated with wound healing and infection, utilizing devices imbedded within dressings or as point-of-care techniques to allow for continual or rapid wound assessment and monitoring. Approaches include using biological or chemical sensors of wound exudates and volatiles to directly or indirectly detect bacteria, monitor pH, temperature, oxygen and enzymes. Spectroscopic and imaging techniques are also reviewed as advanced wound monitoring techniques. The review concludes with a discussion of the limitations of and future directions for this field. Copyright © 2012 Elsevier B.V. All rights reserved.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Molecular pathology of wound healing.

              Skin-wound healing is an orchestrated biological phenomena consisting of three sequential phases, inflammation, proliferation, and maturation. Many biological substances are involved in the process of wound repair, and this short and simplified overview of wound healing can be adopted to determine wound vitality or wound age in forensic medicine. With the development of genetically engineered animals, essential molecules for skin-wound healing have been identified. Especially, cytokines, and growth factors are useful candidates and markers for the determination of wound vitality or age. Moreover, bone marrow-derived progenitor cells would give significant information to wound age determination. In this review article, some interesting observations are presented, possibly contributing to the future practice of forensic pathologists. Copyright © 2010. Published by Elsevier Ireland Ltd.
                Bookmark

                Author and article information

                Journal
                Med Devices (Auckl)
                Med Devices (Auckl)
                Medical Devices: Evidence and Research
                Medical Devices (Auckland, N.Z.)
                Dove Medical Press
                1179-1470
                2018
                15 August 2018
                : 11
                : 267-273
                Affiliations
                [1 ]Minimally Invasive Therapies Group, Medtronic, North Haven, CT, USA, Andrew.m.miesse@ 123456medtronic.com
                [2 ]Minimally Invasive Therapies Group, Medtronic, Plymouth, MN, USA
                [3 ]Minimally Invasive Therapies Group, Medtronic, Mansfield, MA, USA
                [4 ]Image IQ, Cleveland, OH, USA
                Author notes
                Correspondence: Andrew M Miesse, Minimally Invasive Therapies Group, Medtronic, 60 Middletown Avenue, North Haven, CT 06473, USA, Tel +1 203 492 7733, Email Andrew.m.miesse@ 123456medtronic.com
                Article
                mder-11-267
                10.2147/MDER.S171357
                6101738
                30147383
                ddd79523-5967-4dec-b829-7b45aa842861
                © 2018 Eschbach et al. This work is published and licensed by Dove Medical Press Limited

                The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License ( http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.

                History
                Categories
                Original Research

                Biotechnology
                micro-computed tomography,wound healing,tri-staple,image processing,laparoscopic surgery,microvasculature

                Comments

                Comment on this article