      Incidental discovery of circle contact lens by MRI: you can’t scan my poker face, circle contact lens as a potential MRI hazard

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Circle contact lenses, also known as color contact lenses and big eye contact lenses, are a type of cosmetic contact lens. It is not generally known that a circle contact lens usually contains iron oxide and other metals, which means their use during magnetic resonance imaging (MRI) is a potential hazard.

          Case presentation

          We present a rare case of incidental discovery of circle contact lenses by MRI and MRI images of circle lenses in vitro.

          Conclusions

          Circle contact lenses usually contain iron oxide, which is a known source of susceptibility artifact on MRI. Not only radiologists and radiographers but also referring physicians should be familiar with the imaging findings and potential risk of scanning circle contact lenses by MRI.

          Related collections

          Most cited references3

          • Record: found
          • Abstract: not found
          • Article: not found

          Tattoo-induced skin burn during MR imaging.

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Artifacts caused by cosmetics in MR imaging of the head.

            Cosmetics can produce severe distortion of the orbital contents in MR imaging. This distortion, typical of compounds that exhibit a permanent magnetic moment, results from iron oxide in the pigments used to produce dark shades of makeup. In general, the artifact created by cosmetics does not interfere with interpreting routine head scans. However, if the orbital contents are the subject of clinical concern, the images may then be rendered nondiagnostic. The artifact created by the makeup is propagated along the frequency-encoding axis of the images.
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              [Educational-industrial complex research on interactions of MR equipments with contact lenses].

                Bookmark

                Author and article information

                Journal
                BMC Med Imaging
                BMC Med Imaging
                BMC Medical Imaging
                BioMed Central
                1471-2342
                2013
                25 March 2013
                : 13
                : 11
                Affiliations
                [1 ]Department of Diagnostic and Interventional Radiology, Gunma University Hospital, 3-39-22 Showa-machi, Maebashi, Gunma 371-8511, Japan
                Article
                1471-2342-13-11
                10.1186/1471-2342-13-11
                3616822
                23530981
                c62b8e27-3177-48c9-8da0-62aed8fbf72c
                Copyright ©2013 Tokue et al.; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 16 December 2012
                : 21 March 2013
                Categories
                Case Report

                Radiology & Imaging
                circle contact lenses,color contact lenses,magnetic resonance imaging (mri),susceptibility artifact

                Comments

                Comment on this article