      Patient‐specific dose quality assurance of single‐isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          Single‐isocenter multiple brain metastasis stereotactic radiosurgery is an efficient treatment modality increasing in clinical practice. The need to provide accurate, patient‐specific quality assurance (QA) for these plans is met by several options. This study reviews some of these options and explores the use of the Octavius 4D as a solution for patient‐specific plan quality assurance.

          Methods

          The Octavius 4D Modular Phantom (O4D) with the 1000 SRS array was evaluated in this study. The array consists of 977 liquid‐filled ion chambers. The center 5.5 cm × 5.5 cm area has a detector spacing of 2.5 mm. The ability of the O4D to reconstruct three‐dimensional (3D) dose was validated against a 3D gel dosimeter, ion chamber, and film measurements. After validation, 15 patients with 2–11 targets had their plans delivered to the phantom. The criteria used for the gamma calculation was 3%/1 mm. The portion of targets which were measurable by the phantom was countable. The accompanying software compiled the measured doses allowing each target to be counted from the measured dose distribution.

          Results

          Spatial resolution was sufficient to verify the high dose distributions characteristic of SRS. Amongst the 15 patients there were 74 targets. Of the 74 targets, 61 (82%) of them were visible on the measured dose distribution. The average gamma passing rate was 99.3% (with sample standard deviation of 0.68%).

          Conclusions

          The high resolution provided by the O4D with 1000 SRS board insert allows for very high‐resolution measurement. This high resolution in turn can allow for high gamma passing rates. The O4D with the 1000 SRS array is an acceptable method of performing quality assurance for single‐isocenter multiple brain metastasis SRS.

          Related collections

          Most cited references22

          • Record: found
          • Abstract: found
          • Article: not found

          A technique for the quantitative evaluation of dose distributions.

          The commissioning of a three-dimensional treatment planning system requires comparisons of measured and calculated dose distributions. Techniques have been developed to facilitate quantitative comparisons, including superimposed isodoses, dose-difference, and distance-to-agreement (DTA) distributions. The criterion for acceptable calculation performance is generally defined as a tolerance of the dose and DTA in regions of low and high dose gradients, respectively. The dose difference and DTA distributions complement each other in their useful regions. A composite distribution has recently been developed that presents the dose difference in regions that fail both dose-difference and DTA comparison criteria. Although the composite distribution identifies locations where the calculation fails the preselected criteria, no numerical quality measure is provided for display or analysis. A technique is developed to unify dose distribution comparisons using the acceptance criteria. The measure of acceptability is the multidimensional distance between the measurement and calculation points in both the dose and the physical distance, scaled as a fraction of the acceptance criteria. In a space composed of dose and spatial coordinates, the acceptance criteria form an ellipsoid surface, the major axis scales of which are determined by individual acceptance criteria and the center of which is located at the measurement point in question. When the calculated dose distribution surface passes through the ellipsoid, the calculation passes the acceptance test for the measurement point. The minimum radial distance between the measurement point and the calculation points (expressed as a surface in the dose-distance space) is termed the gamma index. Regions where gamma > 1 correspond to locations where the calculation does not meet the acceptance criteria. The determination of gamma throughout the measured dose distribution provides a presentation that quantitatively indicates the calculation accuracy. Examples of a 6 MV beam penumbra are used to illustrate the gamma index.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218.

            Patient-specific IMRT QA measurements are important components of processes designed to identify discrepancies between calculated and delivered radiation doses. Discrepancy tolerance limits are neither well defined nor consistently applied across centers. The AAPM TG-218 report provides a comprehensive review aimed at improving the understanding and consistency of these processes as well as recommendations for methodologies and tolerance limits in patient-specific IMRT QA.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Polymer gel dosimetry.

              Polymer gel dosimeters are fabricated from radiation sensitive chemicals which, upon irradiation, polymerize as a function of the absorbed radiation dose. These gel dosimeters, with the capacity to uniquely record the radiation dose distribution in three-dimensions (3D), have specific advantages when compared to one-dimensional dosimeters, such as ion chambers, and two-dimensional dosimeters, such as film. These advantages are particularly significant in dosimetry situations where steep dose gradients exist such as in intensity-modulated radiation therapy (IMRT) and stereotactic radiosurgery. Polymer gel dosimeters also have specific advantages for brachytherapy dosimetry. Potential dosimetry applications include those for low-energy x-rays, high-linear energy transfer (LET) and proton therapy, radionuclide and boron capture neutron therapy dosimetries. These 3D dosimeters are radiologically soft-tissue equivalent with properties that may be modified depending on the application. The 3D radiation dose distribution in polymer gel dosimeters may be imaged using magnetic resonance imaging (MRI), optical-computerized tomography (optical-CT), x-ray CT or ultrasound. The fundamental science underpinning polymer gel dosimetry is reviewed along with the various evaluation techniques. Clinical dosimetry applications of polymer gel dosimetry are also presented.
                Bookmark

                Author and article information

                Contributors
                SaenzDL@uthscsa.edu
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                18 July 2020
                September 2020
                : 21
                : 9 ( doiID: 10.1002/acm2.v21.9 )
                : 107-115
                Affiliations
                [ 1 ] Department of Radiation Oncology University of Texas Health San Antonio San Antonio TX USA
                Author notes
                [*] [* ] Author to whom correspondence should be addressed. Daniel L. Saenz

                E‐mail: SaenzDL@ 123456uthscsa.edu

                Article
                ACM212979
                10.1002/acm2.12979
                7497916
                c312e8eb-27d3-4546-b0ef-2efda5dd2427
                © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

                This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 23 March 2020
                : 18 June 2020
                : 21 June 2020
                Page count
                Figures: 5, Tables: 4, Pages: 9, Words: 6293
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                September 2020
                Converter:WILEY_ML3GV2_TO_JATSPMC version:5.9.1 mode:remove_FC converted:17.09.2020

                multiple brain metastases,patient specific,quality assurance,single isocenter,stereotactic radiosurgery

                Comments

                Comment on this article