      Converting dose-length product to effective dose at CT.

      Radiology
      Algorithms, Body Burden, Radiation Dosage, Relative Biological Effectiveness, Tomography, X-Ray Computed, Whole-Body Counting, methods

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To determine effective dose (ED) per unit dose-length product (DLP) conversion factors for computed tomographic (CT) dosimetry. A CT dosimetry spreadsheet was used to compute patient ED values and corresponding DLP values. The ratio of ED to DLP was determined with 16-section CT scanners from four vendors, as well as with five models from one manufacturer that spanned more than 25 years. ED-to-DLP ratios were determined for 2-cm scan lengths along the patient axis, as well as for typical scan lengths encountered at head and body CT examinations. The dependence of the ratio of ED to DLP on x-ray tube voltage (in kilovolts) was investigated, and the values obtained with the spreadsheet were compared with those obtained by using two other commercially available CT dosimetry software packages. For 2-cm scan lengths, changes in the scan region resulted in differences to ED of a factor of 30, but much lower variation was obtained for typical scan lengths at clinical head and body imaging. Inter- and intramanufacturer differences for ED/DLP were generally small. Representative values of ED/DLP at 120 kV were 2.2 microSv/mGy x cm (head scans), 5.4 microSv/mGy x cm (cervical spine scans), and 18 microSv/mGy x cm (body scans). For head scans, ED/DLP was approximately independent of x-ray tube voltage, but for body scans, the increase from 80 to 140 kV increased the ratio of ED to DLP by approximately 25%. Agreement in ED/DLP data for all three software packages was generally very good, except for cervical spine examinations where one software package determined an ED/DLP ratio that was approximately double that of the other two. This article describes a method of providing CT users with a practical and reliable estimate of adult patient EDs by using the DLP displayed on the CT console at the end of any given examination. RSNA, 2008

          Related collections

          Author and article information

          Journal
          18710988
          2657852
          10.1148/radiol.2483071964

          Chemistry
          Algorithms,Body Burden,Radiation Dosage,Relative Biological Effectiveness,Tomography, X-Ray Computed,Whole-Body Counting,methods

          Comments

          Comment on this article