      Electrical activity of the diaphragm (EAdi) as a monitoring parameter in difficult weaning from respirator: a pilot study

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction

          A reliable prediction of successful weaning from respiratory support may be crucial for the overall outcome of the critically ill patient. The electrical activity of the diaphragm (EAdi) allows one to monitor the patients’ respiratory drive and their ability to meet the increased respiratory demand. In this pilot study, we compared the EAdi with conventional parameters of weaning failure, such as the ratio of respiratory rate to tidal volume.

          Methods

          We studied 18 mechanically ventilated patients considered difficult to wean. For a spontaneous breathing trial (SBT), the patients were disconnected from the ventilator and given oxygen through a T-piece. The SBT was evaluated by using standard criteria.

          Results

          Twelve patients completed the SBT successfully, and six failed. The EAdi was significantly different in the two groups. We found an early increase in EAdi in the failing patients that was more pronounced than in any of the patients who successfully passed the SBT. Changes in EAdi predicted an SBT failure earlier than did conventional parameters.

          Conclusions

          EAdi monitoring adds valuable information during weaning from the ventilator and may help to identify patients who are not ready for discontinuation of respiratory support.

          Related collections

          Most cited references35

          • Record: found
          • Abstract: found
          • Article: not found

          Weaning from mechanical ventilation.

          Weaning covers the entire process of liberating the patient from mechanical support and from the endotracheal tube. Many controversial questions remain concerning the best methods for conducting this process. An International Consensus Conference was held in April 2005 to provide recommendations regarding the management of this process. An 11-member international jury answered five pre-defined questions. 1) What is known about the epidemiology of weaning problems? 2) What is the pathophysiology of weaning failure? 3) What is the usual process of initial weaning from the ventilator? 4) Is there a role for different ventilator modes in more difficult weaning? 5) How should patients with prolonged weaning failure be managed? The main recommendations were as follows. 1) Patients should be categorised into three groups based on the difficulty and duration of the weaning process. 2) Weaning should be considered as early as possible. 3) A spontaneous breathing trial is the major diagnostic test to determine whether patients can be successfully extubated. 4) The initial trial should last 30 min and consist of either T-tube breathing or low levels of pressure support. 5) Pressure support or assist-control ventilation modes should be favoured in patients failing an initial trial/trials. 6) Noninvasive ventilation techniques should be considered in selected patients to shorten the duration of intubation but should not be routinely used as a tool for extubation failure.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Evidence-based guidelines for weaning and discontinuing ventilatory support: a collective task force facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care Medicine.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Extubation outcome after spontaneous breathing trials with T-tube or pressure support ventilation. The Spanish Lung Failure Collaborative Group.

              A 2-h T-tube trial of spontaneous breathing was used in selecting patients ready for extubation and discontinuation of mechanical ventilation. However, some doubt remains as to whether it is the most appropriate method of performing a spontaneous breathing trial. We carried out a prospective, randomized, multicenter study involving patients who had received mechanical ventilation for more than 48 h and who were considered by their physicians to be ready for weaning according to clinical criteria and standard weaning parameters. Patients were randomly assigned to undergo a 2-h trial of spontaneous breathing in one of two ways: with a T-tube system or with pressure support ventilation of 7 cm H2O. If a patient had signs of poor tolerance at any time during the trial, mechanical ventilation was reinstituted. Patients without these features at the end of the trial were extubated. Of the 246 patients assigned to the T-tube group, 192 successfully completed the trial and were extubated; 36 of them required reintubation. Of the 238 patients in the group receiving pressure support ventilation, 205 were extubated and 38 of them required reintubation. The percentage of patients who remained extubated after 48 h was not different between the two groups (63% T-tube, 70% pressure support ventilation, p = 0.14). The percentage of patients falling the trial was significantly higher when the T-tube was used (22 versus 14%, p = 0.03). Clinical evolution during the trial was not different in patients reintubated and successfully extubated. ICU mortality among reintubated patients was significantly higher than in successfully extubated patients (27 versus 2.6%, p < 0.001). Spontaneous breathing trials with pressure support or T-tube are suitable methods for successful discontinuation of ventilator support in patients without problems to resume spontaneous breathing.
                Bookmark

                Author and article information

                Contributors
                Journal
                Crit Care
                Crit Care
                Critical Care
                BioMed Central
                1364-8535
                1466-609X
                2013
                28 August 2013
                : 17
                : 4
                : R182
                Affiliations
                [1 ]Department of Anesthesiology, Emergency and Intensive Care Medicine, University of Göttingen Medical School (Georg-August University of Goettingen), Robert-Koch-Str. 40, D-37099 Göttingen, Germany
                [2 ]Scuola di Specializzazione in Anestesia, Rianimazione e Terapia Antalgica, Università degli Studi di Pavia, IRCCS Policlinico S.Matteo, Viale Camillo Golgi, 19, 27100 Pavia, Province of Pavia, Italy
                [3 ]Dipartimento di Anestesia e Rianimazione 1°, Ospedale S. Bortolo di Vicenza, viale Rodolfi, 37-36100 Vicenza, Italy
                Article
                cc12865
                10.1186/cc12865
                4057029
                23985299
                7419e0f5-1a5a-4536-9468-20d92e73fb06
                Copyright © 2013 Barwing et al.; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 24 January 2013
                : 2 August 2013
                Categories
                Research

                Emergency medicine & Trauma
                electrical activity of the diaphragm (eadi),neurally adjusted ventilatory assist (nava),ventilator weaning

                Comments

                Comment on this article