      Dipstick measurements of urine specific gravity are unreliable.

      Archives of Disease in Childhood
      Humans, Hydrogen-Ion Concentration, Osmolar Concentration, Point-of-Care Systems, standards, Reagent Kits, Diagnostic, Refractometry, Reproducibility of Results, Specific Gravity, Urinalysis, methods

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate the reliability of dipstick measurements of urine specific gravity (U-SG). Fresh urine specimens were tested for urine pH and osmolality (U-pH, U-Osm) by a pH meter and an osmometer, and for U-SG by three different methods (refractometry, automatic readout of a dipstick (Clinitek-50), and (visual) change of colour of the dipstick). The correlations between the visual U-SG dipstick measurements and U-SG determined by a refractometer and the comparison of Clinitek((R))-50 dipstick U-SG measurements with U-Osm were less than optimal, showing very wide scatter of values. Only the U-SG refractometer values and U-Osm had a good linear correlation. The tested dipstick was unreliable for the bedside determination of U-SG, even after correction for U-pH, as recommended by the manufacturer. Among the bedside determinations, only refractometry gives reliable U-SG results. Dipstick U-SG measurements should be abandoned.