      Clinical application of a GPU-accelerated monte carlo dose verification for cyberknife M6 with Iris collimator

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          To apply an independent GPU-accelerated Monte Carlo (MC) dose verification for CyberKnife M6 with Iris collimator and evaluate the dose calculation accuracy of RayTracing (TPS-RT) algorithm and Monte Carlo (TPS-MC) algorithm in the Precision treatment planning system (TPS).

          Methods

          GPU-accelerated MC algorithm (ArcherQA-CK) was integrated into a commercial dose verification system, ArcherQA, to implement the patient-specific quality assurance in the CyberKnife M6 system. 30 clinical cases (10 cases in head, and 10 cases in chest, and 10 cases in abdomen) were collected in this study. For each case, three different dose calculation methods (TPS-MC, TPS-RT and ArcherQA-CK) were implemented based on the same treatment plan and compared with each other. For evaluation, the 3D global gamma analysis and dose parameters of the target volume and organs at risk (OARs) were analyzed comparatively.

          Results

          For gamma pass rates at the criterion of 2%/2 mm, the results were over 98.0% for TPS-MC vs.TPS-RT, TPS-MC vs. ArcherQA-CK and TPS-RT vs. ArcherQA-CK in head cases, 84.9% for TPS-MC vs.TPS-RT, 98.0% for TPS-MC vs. ArcherQA-CK and 83.3% for TPS-RT vs. ArcherQA-CK in chest cases, 98.2% for TPS-MC vs.TPS-RT, 99.4% for TPS-MC vs. ArcherQA-CK and 94.5% for TPS-RT vs. ArcherQA-CK in abdomen cases. For dose parameters of planning target volume (PTV) in chest cases, the deviations of TPS-RT vs. TPS-MC and ArcherQA-CK vs. TPS-MC had significant difference ( P < 0.01), and the deviations of TPS-RT vs. TPS-MC and TPS-RT vs. ArcherQA-CK were similar ( P > 0.05). ArcherQA-CK had less calculation time compared with TPS-MC (1.66 min vs. 65.11 min).

          Conclusions

          Our proposed MC dose engine (ArcherQA-CK) has a high degree of consistency with the Precision TPS-MC algorithm, which can quickly identify the calculation errors of TPS-RT algorithm for some chest cases. ArcherQA-CK can provide accurate patient-specific quality assurance in clinical practice.

          Related collections

          Most cited references26

          • Record: found
          • Abstract: found
          • Article: not found

          BEAM: a Monte Carlo code to simulate radiotherapy treatment units.

          This paper describes BEAM, a general purpose Monte Carlo code to simulate the radiation beams from radiotherapy units including high-energy electron and photon beams, 60Co beams and orthovoltage units. The code handles a variety of elementary geometric entities which the user puts together as needed (jaws, applicators, stacked cones, mirrors, etc.), thus allowing simulation of a wide variety of accelerators. The code is not restricted to cylindrical symmetry. It incorporates a variety of powerful variance reduction techniques such as range rejection, bremsstrahlung splitting and forcing photon interactions. The code allows direct calculation of charge in the monitor ion chamber. It has the capability of keeping track of each particle's history and using this information to score separate dose components (e.g., to determine the dose from electrons scattering off the applicator). The paper presents a variety of calculated results to demonstrate the code's capabilities. The calculated dose distributions in a water phantom irradiated by electron beams from the NRC 35 MeV research accelerator, a Varian Clinac 2100C, a Philips SL75-20, an AECL Therac 20 and a Scanditronix MM50 are all shown to be in good agreement with measurements at the 2 to 3% level. Eighteen electron spectra from four different commercial accelerators are presented and various aspects of the electron beams from a Clinac 2100C are discussed. Timing requirements and selection of parameters for the Monte Carlo calculations are discussed.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Report of AAPM TG 135: quality assurance for robotic radiosurgery.

            The task group (TG) for quality assurance for robotic radiosurgery was formed by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance (QA) Subcommittee. The task group (TG-135) had three main charges: (1) To make recommendations on a code of practice for Robotic Radiosurgery QA; (2) To make recommendations on quality assurance and dosimetric verification techniques, especially in regard to real-time respiratory motion tracking software; (3) To make recommendations on issues which require further research and development. This report provides a general functional overview of the only clinically implemented robotic radiosurgery device, the CyberKnife. This report includes sections on device components and their individual component QA recommendations, followed by a section on the QA requirements for integrated systems. Examples of checklists for daily, monthly, annual, and upgrade QA are given as guidance for medical physicists. Areas in which QA procedures are still under development are discussed.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found
              Is Open Access

              Stereotactic radiosurgery in the treatment of brain metastases: the current evidence.

              Chemotherapy has made substantial progress in the therapy of systemic cancer, but the pharmacological efficacy is insufficient in the treatment of brain metastases. Fractionated whole brain radiotherapy (WBRT) has been a standard treatment of brain metastases, but provides limited local tumor control and often unsatisfactory clinical results. Stereotactic radiosurgery using Gamma Knife, Linac or Cyberknife has overcome several of these limitations, which has influenced recent treatment recommendations. This present review summarizes the current literature of single session radiosurgery concerning survival and quality of life, specific responses, tumor volumes and numbers, about potential treatment combinations and radioresistant metastases. Gamma Knife and Linac based radiosurgery provide consistent results with a reproducible local tumor control in both single and multiple brain metastases. Ideally minimum doses of ≥18Gy are applied. Reported local control rates were 90-94% for breast cancer metastases and 81-98% for brain metastases of lung cancer. Local tumor control rates after radiosurgery of otherwise radioresistant brain metastases were 73-90% for melanoma and 83-96% for renal cell cancer. Currently, there is a tendency to treat a larger number of brain metastases in a single radiosurgical session, since numerous studies document high local tumor control after radiosurgical treatment of >3 brain metastases. New remote brain metastases are reported in 33-42% after WBRT and in 39-52% after radiosurgery, but while WBRT is generally applied only once, radiosurgery can be used repeatedly for remote recurrences or new metastases after WBRT. Larger metastases (>8-10cc) should be removed surgically, but for smaller metastases Gamma Knife radiosurgery appears to be equally effective as surgical tumor resection (level I evidence). Radiosurgery avoids the impairments in cognition and quality of life that can be a consequence of WBRT (level I evidence). High local efficacy, preservation of cerebral functions, short hospitalization and the option to continue a systemic chemotherapy are factors in favor of a minimally invasive approach with stereotactic radiosurgery.
                Bookmark

                Author and article information

                Contributors
                sr_qren@163.com
                sinkriver@126.com
                Journal
                Radiat Oncol
                Radiat Oncol
                Radiation Oncology (London, England)
                BioMed Central (London )
                1748-717X
                2 July 2024
                2 July 2024
                2024
                : 19
                : 86
                Affiliations
                [1 ]GRID grid.410570.7, ISNI 0000 0004 1760 6682, Department of Cancer Center, Daping Hospital, , Army Medical University, ; Chongqing, China
                [2 ]School of Nuclear Science and Technology, University of Science and Technology of China, ( https://ror.org/04c4dkn09) Hefei, China
                [3 ]Department of Radiation Oncology, Dezhou Second People’s Hospital, ( https://ror.org/042g3qa69) Dezhou, China
                [4 ]Anhui Wisdom Technology Company Limited, Hefei, China
                Article
                2446
                10.1186/s13014-024-02446-1
                11221037
                38956685
                624fb859-e2ee-43a6-abf7-b0e88a91bf32
                © The Author(s) 2024

                Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

                History
                : 8 October 2023
                : 29 April 2024
                Funding
                Funded by: Chongqing medical scientific research project (Joint project of Chongqing Health Commission and Science and Technology Bureau)
                Award ID: 2022ZDXM027
                Award ID: 2022ZDXM027
                Award ID: 2022ZDXM027
                Award ID: 2022ZDXM027
                Categories
                Research
                Custom metadata
                © BioMed Central Ltd., part of Springer Nature 2024

                Oncology & Radiotherapy
                gpu-accelerated monte carlo,cyberknife,patient-specific quality assurance

                Comments

                Comment on this article