      Technical note on introducing a digital workflow for newborns with craniofacial anomalies based on intraoral scans - part II: 3D printed Tübingen palatal plate prototype for newborns with Robin sequence

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Orthodontic treatment of newborns and infants with Robin-Sequence using the Tübingen Palatal Plate (TPP) is a complex procedure that could benefit from simplification through digitalization. The design of the velar extension (spur) and the palatal base determines the success of the treatment. Therefore, a prototype must be produced and inserted under endoscopic supervision in order to determine the appropriate shape, length and position of the spur. This technical note demonstrates a fully digital workflow for the design and manufacturing of a functional TPP prototype, based on an intraoral scan. This prototype can be altered and individualized digitally for each patient. After the shape and position of the spur have been optimized, the prototype is duplicated using a silicone mold. Then the definitive TPP is manufactured and inserted. We aim to present a workflow which facilitates the fitting procedure and does not require a conventional impression or a physical model to create the appliance.

          Methods

          As described in part I of this series, the intraoral scan is performed using the 3Shape TRIOS3 scanner and its corresponding acquisition software. The virtual model is rendered in the 3Shape ortho appliance designer and the base of the palatal plate is designed in the 3Shape dental designer. The palatal plate and the virtual model are then imported into Autodesk Meshmixer and a standardized spur is positioned and merged with the base. The TPP is exported in Standard Tessellation Language (STL) format and manufactured on a W2P Solflex 170 DLP printer using VOCO VPrint Splint material (MDR Class IIa).

          Results

          Based on an intraoral scan, the TPP prototype could be successfully manufactured and proved suitable for the patients’ treatment.

          Conclusion

          The new digital workflow for the design of the TPP can been successfully implemented into daily clinical routine in our facility. Patients could be alleviated from having to undergo conventional impression procedures and fitting of the TPP could be facilitated by producing multiple functional prototypes for endoscopic evaluation. Through rapid prototyping, the expenditure of the fitting process was reduced, which makes the TPP therapy more efficient and accessible to a wider range of clinicians.

          Related collections

          Most cited references7

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Conventional Vs Digital Impressions: Acceptability, Treatment Comfort and Stress Among Young Orthodontic Patients

          Objective: The objective of the present study was to compare patients’ acceptability, comfort and stress with conventional and digital impressions. Materials and Methods: Thirty young orthodontic patients (15 males and 15 females) who had no previous experience of impressions were enrolled in this study. Conventional impressions for orthodontic study models of the dental arches were taken using an alginate impression material (Hydrogum®, Zhermack Spa, Badia Polesine, Rovigo, Italy). Fifteen days later, digital impressions of both arches were acquired using an intraoral scanner (CS3600®, Carestream Dental, Rochester, NY, USA). Immediately after impression taking, patients’ acceptability, comfort and stress were measured using two questionnaires and the State anxiety scale. Results: Data showed no difference in terms of anxiety and stress; however, patients preferred the use of digital impressions systems instead of conventional impression techniques. Alginate impressions resulted as fast as digital impressions. Conclusions: Digital impressions resulted the most accepted and comfortable impression technique in young orthodontic patients, when compared to conventional techniques.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Intraoral 3D Scanning or Dental Impressions for the Assessment of Dental Arch Relationships in Cleft Care: Which is Superior?

            This study was undertaken to evaluate intraoral 3D scans for assessing dental arch relationships and obtain patient/parent perceptions of impressions and intraoral 3D scanning.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Technical note on introducing a digital workflow for newborns with craniofacial anomalies based on intraoral scans - part I: 3D printed and milled palatal stimulation plate for trisomy 21

              Background Advanced digital workflows in orthodontics and dentistry often require a combination of different software solutions to create patient appliances, which may be a complex and time-consuming process. The main objective of this technical note is to discuss treatment of craniofacial anomalies using digital technologies. We present a fully digital, linear workflow for manufacturing palatal plates for infants with craniofacial anomalies based on intraoral scanning. Switching to intraoral scanning in infant care is advantageous as taking conventional impressions carries the risk of impression material aspiration and/or infections caused by material remaining in the oronasal cavity. Material and methods The fully digital linear workflow presented in this technical note can be used to design and manufacture palatal plates for cleft palate patients as well as infants with functional disorders. We describe the workflow implemented in an infant with trisomy 21. The maxilla was registered using a digital scanner and a stimulation plate was created using dental CAD software and an individual impression tray module on a virtual model. Plates were manufactured using both additive and subtractive methods. Methacrylate based light curing resin and Poly-Ether-Ether-Ketone were the materials used. Results The palatal area was successfully scanned to create a virtual model. The plates fitted well onto the palatal area. Manual post-processing was necessary to optimize a functional ridge along the vestibular fold and remove support structures from the additively manufactured plate as well as the milled plate produced from a blank. The additively manufactured plate fitted better than the milled one. Conclusion Implementing a fully digital linear workflow into clinical routine for treatment of neonates and infants with craniofacial disorders is feasible. The software solution presented here is suitable for this purpose and does not require additional software for the design. This is the key advantage of this workflow, which makes digital treatment accessible to all clinicians who want to deal with digital technology. Whether additive or subtractive manufacturing is preferred depends on the appliance material of choice and influences the fit of the appliance.
                Bookmark

                Author and article information

                Contributors
                alexander.xepapadeas@med.uni-tuebingen.de
                Journal
                BMC Oral Health
                BMC Oral Health
                BMC Oral Health
                BioMed Central (London )
                1472-6831
                16 June 2020
                16 June 2020
                2020
                : 20
                : 171
                Affiliations
                [1 ]GRID grid.411544.1, ISNI 0000 0001 0196 8249, Department of Orthodontics, , University Hospital Tuebingen, ; Osianderstr. 2-8, 72076 Tuebingen, Germany
                [2 ]GRID grid.411544.1, ISNI 0000 0001 0196 8249, Section “Medical Materials Science & Technology” University Hospital Tuebingen, ; Osianderstr. 2-8, 72076 Tuebingen, Germany
                [3 ]GRID grid.411544.1, ISNI 0000 0001 0196 8249, Department of Neonatology, , University Hospital Tuebingen, ; Calwerstr. 7, 72076 Tuebingen, Germany
                Author information
                http://orcid.org/0000-0003-1645-929X
                Article
                1159
                10.1186/s12903-020-01159-7
                7298740
                32546229
                22b6f57d-aaec-412b-af68-228e6d51431f
                © The Author(s) 2020

                Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

                History
                : 6 October 2019
                : 4 June 2020
                Categories
                Technical Advance
                Custom metadata
                © The Author(s) 2020

                Dentistry
                additive manufacturing (am),digital workflow,vat-polymerization,cleft lip and palate (clp),down’s syndrome (ds),robin-sequence (rs),tübingen palatal plate (tpp),computer-aided design/computer-aided manufacturing (cad/cam),computer-aided impression (cai),functional prototyping

                Comments

                Comment on this article

                scite_
                0
                0
                0
                0
                Smart Citations
                0
                0
                0
                0
                Citing PublicationsSupportingMentioningContrasting
                View Citations

                See how this article has been cited at scite.ai

                scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

                Similar content111

                Cited by14

                Most referenced authors56