      Monte Carlo Modeling of the Agility MLC for IMRT and VMAT Calculations

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background/Aim: The Purpose of this study was to develop a Monte Carlo (MC) model for the Agility multileaf collimator (MLC) mounted and to validate its accuracy. Materials and Methods: To describe the Agility MLC in the BEAMnrc MC code, an existing component module code was modified to include its characteristics. The leaf characterization of the MC model was validated by comparing the calculated interleaf transmission and tongue-and-groove effect with EBT2 film and diode measurements and IMRT and VMAT calculations with film measurements. Results: Agreement between mean calculated and measured leaf transmissions was within 0.1%. The discrepancy between MC calculation and measurement in a static irregular field was less than 2%/2 mm. Gamma analysis of the comparison of MC and EBT2 film measurements in IMRT and VMAT fields yielded pass rates of 99.1% and 99.5% with 3%/3 mm criteria, respectively. Conclusion: Our findings demonstrate the accuracy of the MC model using an adapted BEAMnrc component module for the Agility MLC.

          Related collections

          Author and article information

          Journal
          In Vivo
          In Vivo
          In Vivo
          International Institute of Anticancer Research
          0258-851X
          1791-7549
          3 September 2020
          Sep-Oct 2020
          : 34
          : 5
          : 2371-2380
          Affiliations
          [1 ]Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan
          [2 ]Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Osaka, Japan
          [3 ]Department of Radiation Oncology, Kansai Medical University Hospital, Osaka, Japan
          Author notes
          Shingo Ohira, Ph.D., Department of Radiation Oncology, Osaka International Cancer Institute, 3-1-69 Otemae, Chuo-ku, Osaka, 537-8567, Japan. Tel: +81 669451181, Fax: +81 669451900 oohira-si@ 123456mc.pref.osaka.jp
          Article
          PMC7652464 PMC7652464 7652464
          10.21873/invivo.12050
          7652464
          32871762
          20447520-62ca-410d-915a-ec8ae9b8b979
          Copyright 2020, International Institute of Anticancer Research
          History
          : 27 May 2020
          : 22 June 2020
          Categories
          Research Article

          QA,Agility MLC,VMAT,IMRT,Monte Carlo
          QA, Agility MLC, VMAT, IMRT, Monte Carlo

          Comments

          Comment on this article