      Technical Report: Reference photon dosimetry data for Varian accelerators based on IROC-Houston site visit data

      other

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose:

          Accurate data regarding linear accelerator (Linac) radiation characteristics are important for treatment planning system modeling as well as regular quality assurance of the machine. The Imaging and Radiation Oncology Core-Houston (IROC-H) has measured the dosimetric characteristics of numerous machines through their on-site dosimetry review protocols. Photon data are presented and can be used as a secondary check of acquired values, as a means to verify commissioning a new machine, or in preparation for an IROC-H site visit.

          Methods:

          Photon data from IROC-H on-site reviews from 2000 to 2014 were compiled and analyzed. Specifically, data from approximately 500 Varian machines were analyzed. Each dataset consisted of point measurements of several dosimetric parameters at various locations in a water phantom to assess the percentage depth dose, jaw output factors, multileaf collimator small field output factors, off-axis factors, and wedge factors. The data were analyzed by energy and parameter, with similarly performing machine models being assimilated into classes. Common statistical metrics are presented for each machine class. Measurement data were compared against other reference data where applicable.

          Results:

          Distributions of the parameter data were shown to be robust and derive from a student’s t distribution. Based on statistical and clinical criteria, all machine models were able to be classified into two or three classes for each energy, except for 6 MV for which there were eight classes. Quantitative analysis of the measurements for 6, 10, 15, and 18 MV photon beams is presented for each parameter; supplementary material has also been made available which contains further statistical information.

          Conclusions:

          IROC-H has collected numerous data on Varian Linacs and the results of photon measurements from the past 15 years are presented. The data can be used as a comparison check of a physicist’s acquired values. Acquired values that are well outside the expected distribution should be verified by the physicist to identify whether the measurements are valid. Comparison of values to this reference data provides a redundant check to help prevent gross dosimetric treatment errors.

          Related collections

          Author and article information

          Journal
          Med Phys
          Med Phys
          MPHYA6
          Medical Physics
          American Association of Physicists in Medicine
          0094-2405
          May 2016
          15 April 2016
          : 43
          : 5
          : 2374-2386
          Affiliations
          Department of Radiation Physics, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030; Imaging and Radiation Oncology Core-Houston, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030; and Graduate School of Biomedical Sciences, The University of Texas Health Science Center-Houston , Houston, Texas 77030
          Department of Radiation Physics, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030 and Imaging and Radiation Oncology Core-Houston, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030
          Department of Biostatistics, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030
          Department of Radiation Physics, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030; Imaging and Radiation Oncology Core-Houston, The University of Texas MD Anderson Cancer Center , Houston, Texas 77030; and Graduate School of Biomedical Sciences, The University of Texas Health Science Center-Houston , Houston, Texas 77030
          Author notes
          [a)]

          Author to whom correspondence should be addressed. Electronic mail: sfkry@ 123456mdanderson.org

          Article
          PMC4833752 PMC4833752 4833752 039605MPH 1.4945697 15-1512R2
          10.1118/1.4945697
          4833752
          27147349
          1d12528e-5b0e-42cb-ba9f-967ef9d44f8b
          © 2016 American Association of Physicists in Medicine.

          0094-2405/2016/43(5)/2374/13/ $30.00

          History
          : 29 September 2015
          : 21 March 2016
          : 27 March 2016
          Page count
          Pages: 13
          Funding
          Funded by: National Cancer Institute (NCI) http://dx.doi.org/10.13039/100000054
          Award ID: CA180803
          Categories
          COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY
          Technical Notes
          Custom metadata

          Linac,reference data,percentage depth dose,small field,IROC
          Linac, reference data, percentage depth dose, small field, IROC

          Comments

          Comment on this article

          scite_
          36
          7
          54
          0
          Smart Citations
          36
          7
          54
          0
          Citing PublicationsSupportingMentioningContrasting
          View Citations

          See how this article has been cited at scite.ai

          scite shows how a scientific paper has been cited by providing the context of the citation, a classification describing whether it supports, mentions, or contrasts the cited claim, and a label indicating in which section the citation was made.

          Similar content256

          Cited by11